Appendix A - Product Configurations; Table 7 - Non-Us Configuration Parameters; Table 8 - Us Configuration Parameters - Cardinal Health Infant Flow SiPAP Operator's Manual

Table of Contents

Advertisement

Operator's Manual

Appendix A - Product Configurations

Non-US Configuration Parameters
Table 7 – Non-US Configuration Parameters
Parameter
Set Oxygen concentration, %O
NCPAP / Pres Low flow rate
Pres High flow rate
BiPhasic / BiPhasic tr* on time, T-
High
BiPhasic rate, R (mandatory rate)
BiPhasic tr* backup rate, Rb (apnea
backup rate)
Apnea timeout, Tapnea
Note
BiPhasic tr mode not currently available in the United States. In non-US
configurations, T-High automatically reduces at higher R and Rb rate settings to
maintain a minimum off time of 100 milliseconds
US Configuration Parameters
Table 8 – US Configuration Parameters
Parameter
Set Oxygen concentration, %O
NCPAP / Pres Low flow rate
Pres High flow rate
BiPhasic on time, T-High
BiPhasic rate, R (Mandatory rate)
Low Breath Rate timeout, T
Note
In US configurations, T-High automatically reduces at higher R and Rb rate settings
to maintain a minimum off time of 1.0 seconds.
675-101-101
Revision H
Min
Max
21
100
2
0
15
0
5
0.1
3.0
1
120
1
120
10
30
Min
Max
21
100
2
0
15
0
5
0.1
3.0 *
1
54
10
30
LBR
Accuracy
Units
±3
%
±15%
L/min
± 15%
L/min
± 0.005
seconds
± 0.5
bpm
± 0.5
bpm
± 1
seconds
.
Accuracy
Units
±3
%
±15%
L/min
± 15%
L/min
± 0.005
seconds
± 0.5
bpm
± 1
Seconds
49
Default
N/A
N/A
N/A
.3 sec
30 bpm
10 bpm
20 sec
Default
N/A
N/A
N/A
1.0 sec
10 bpm
20 sec