Locking The Device Positions - Agilent Technologies BenchCel User Manual

Microplate handler
Table of Contents

Advertisement

4 Setting up the BenchCel device
Setting and managing teachpoints

Locking the device positions

C AU T I O N
90
Tightening the clamp screw can cause the teachpoint to shift. Make sure you verify
the teachpoint coordinates after locking the device position.
Before you record the teachpoint, lock the devices in their positions.
To lock the devices in their positions:
1
Tighten the BenchCel integration plate clamp screw using the 3-mm hex wrench.
2
Check the teachpoint coordinates again. See
coordinates" on page
"Determining the teachpoint
86.
Tighten the screw to lock
the devices in position.
BenchCel Microplate Handler User Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BenchCel and is the answer not in the manual?

Questions and answers

Table of Contents