7 Operating the bed

The birthing bed is not designed for permanent operation – motors can only be loaded up to
the values defined on the label, i.e. 2 minutes of operation followed by 18 minutes of stand
by. Excess workload of the device may result in automatic stop. Under such conditions,
authorized service action is needed.
The bed includes firmware that can be updated only by an authorised service technician.
This firmware is protected against unauthorised access by mechanical housing (tool is
needed to access), by seal (components with processor are sealed), by exclusive
compatibility with an authorised software tool and by check of compatibility of the new
firmware with the bed.
The bed includes firmware that can be updated only by an authorised service technician.
This firmware is protected against unauthorised access by mechanical housing (tool is
needed to access), by seal (components with processor are sealed), by exclusive
compatibility with an authorised software tool and by check of compatibility of the new
firmware with the bed.
Components of the product are not treated against fire. Hence, avoid handling near
open fire. If this rule is not observed, the device may catch fire.
When handling the product, please pay attention to t h e protection of a power cable
against mechanical damage, especially its insulation.
In case of fuse replacement pay attention to correct selection of a fuse type. Fuses
need not be replaced by an authorized BORCAD Medical a.s. service; it can be done by
a qualified electrician.
Pay attention to free passage of hoses, cannulae, and other free parts (in case infusions
are used) so as not to get these pinched in the moving parts of the product (rails or
leaning parts may be the risk locations).
When positioning the product, collision with objects may occur. Keeping safe distance
from other objects is recommended in order to prevent damage and injury.

7.1 Side rails

The woman sitting in the chair can use the handles for increasing her sitting stability
or pressing internal functional buttons to increase her sitting comfort.
Side rails are also intended for medical professionals. During closer access to
woman in labor siderials can be parked away from outer side.

7.1.1 Description of the control panel for the woman in childbirth

BORCAD Medical a.s.
739 45 Fryčovice 673, tel.: +420 558 640 631
Warning:
Caution
Lowering / lifting of
back section
Height
adjustment