9 Care

All BORCAD Medical a.s. products are designed for many years of service under
the conditions of correct use. Please use your bed and its accessories carefully.
Please follow these instructions for use and recommended operation and
maintenance instructions. Avoid negligent handling. Do not, in case of a
function failure, undertake any repairs and never use force!

9.1 Cleaning and disinfection

Do not use any automated cleaning or disinfection methods!
For basic cleaning use a soft cloth and only a pH-neutral detergent solution of an
multi-purpose cleaner.
For disinfection, we recommend using standard preparations for disinfection of
surfaces based on for example quaternary ammonia compounds, peroxy-
compounds. The dilution ratio recommended by the Manufacturer of these products
must be followed at all times!
Do not use any abrasive washing aids or ketone based solvents.
By using too much cleaner you risk water leakage into the motor parts of the
product, which may damage the electric wiring and render the product not functional.
In case of improperly performed cleaning and disinfection, there is a risk of
nosocomial infection.
Put the upholstery on dry bed areas only.
When cleaning, place the footrest upholstery on the flat side.
This medical device with all its parts and accessories is not designed to undergo
sterilization.
BORCAD Medical a.s. is not responsible for damage to the cushioning (e.g.
cracking, paint flaking etc.), which would result from improper use of disinfectants.
In such cases, warranty claim shall not apply.
According to the principles of good hygienic practice, the cleaning and disinfection of
the delivery room, its equipment and all the medical devices used, must be made
directly and promptly after the delivery (birth) in order to prepare the equipment for a
new delivery. These principles apply also to the cleaning and disinfection of the
delivery bed. All parts of the delivery bed must be cleaned. The procedure must be
verified and/or supplemented to be in compliance with the respective hospital
BORCAD Medical a.s.