7.7 Backup Battery

The batteries may be only replaced by the appropriate service personnel!
The battery is intended solely for emergency power supply during power failures.
In case that the insufficiently trained personnel will perform the replacement of battery, the
danger situation may occur or the functionality of medical device may be reduced.
Backup battery discharge is indicated by sound signal (1 s) after depressing any movement
button. In this case, connect the bed to the mains immediately and charge the battery. The
bed should not be used, when the battery has been completely discharged.
When the back-up battery is completely discharged, the recharging to the minimum level
(allowing the use of this back-up battery again) will take at least one hour time If you require
the full charging of the battery, the necessary time for charging will take at least 24 hours!
The fully charged back-up battery will last in the stand-by mode for approxi-mately 1 to 2
months (at the optimum temperature of 15 to 25°C). The allowed maximum temperature for
the battery is 50°C!
The fully charged battery provides for the necessary functionality of the birthing bed (i.e.
positioning of lift, tilting of the supporting systems, calling the nurse, and Trendelenburg),
namely with minimum of four cycles, so that it will be possible to set the birthing bed into the
basic safe position.
In case that the power supply of the bed will be interrupted, the "Call nurse" function will be
usable for the time period of 1 to 2 months (provided that the positioning of the bed is not
used!).
The birthing bed includes fully charged battery. The battery is automatically charged
from the control unit and does not require special maintenance.
Indication of battery charging mode
Battery charging mode is indicated by a green indicator LED on side control panel.
Battery status indicator
Light on: battery disconnected or defective
Flashing: charging the battery
Light off: charging completed