Change Method Parameters; Method Type; Titration Mode - Xylem TitroLine 7800 Operating Manual

Table of Contents

Advertisement

4.6

Change Method Parameters

The input or modification of the method name was already described in  4.1 and 4.3
Fig. 98

Method type

4.6.1
On the «Method type» you can select whether you wish to perform a manual or automatic titration, a dosage or
whether you wish to prepare a solution.
Fig. 99
The selection of the Method type will have an influence on the further parameterisation of the method
For instance, if you select the dosing mode, neither a selection of a formula nor a change of the automatic
titration mode (KF and dead stop) will be available.

Titration mode

4.6.2
For an automatic titration, you can select from the following modes:
Linear titration (pH and mV)
Dynamic titration (pH and mV)
End-Point titration (pH, mV and µA)
Dead Stop titration (µA)
KF- Titration (see  4.7 Method parameters of the KF-Titration)
pH Stat Titration (pH)
In addition one can also carry out a measurement: