Xylem TitroLine 7800 Operating Manual page 93

Table of Contents

Advertisement

4.7.2.7 Titration end
The end of a titration (Fig. 195 and Fig. 196) is reached, and the result will be calculated as soon as, or if,
respectively:
The defined End value in µA value has been reached
The Endpoint delay in seconds has been adhered
The drift value in µg/min has been reached
The predefined value ml has been reached (Maximum titration volume)
The conditions for minimum and maximum titration time are maintained
Fig. 195
Fig. 196
Maximum titration time
Can be set between 0 - 9999 seconds.
The default setting is 600 seconds. The maximum titration time is generally used for KF titration, which can
create a high continuous drift from a secondary reaction and thus cannot reach a stable endpoint.
Minimum titration time
Can be set between 0 - 9999 seconds. The default setting is 10 seconds.
The minimum titration time prevents premature termination of the titration if there is a delay in the extraction of
water from the sample. The minimum titration time is used in combination with the extraction time. It expires while
the extraction time is still active.
237

Advertisement

Table of Contents
loading

Table of Contents