Table of Contents

Advertisement

185

KF Titration

3.6.4
The method being displayed can now be carried out immediately with <START>.
The preconditioning is run first.
The solvent and the titration vessel contain moisture (water) that should not influence the calculation of the result.
The conditioning is run automatically after pressing <START> (Fig. 69). The final conditions are the same as the
conditions of the actual sample titration.
Fig. 69
When the final criteria are met, then there is an audible signal and
a message appears
(Fig.
70).
Fig. 70
The conditioning remains active until the actual titration is started by pressing <Start>.
You are prompted immediately to add the sample (Fig. 71).
Fig. 71

Advertisement

Table of Contents
loading

Table of Contents