Technical Specifications; Titrator Titroline ® 7800 - Xylem TitroLine 7800 Operating Manual

Table of Contents

Advertisement

150
1.3

Technical Specifications

Titrator TitroLine
1.3.1
Translation of the legally binding German version
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
 All common titration solutions.
 As reagent water and all non-aggressive non-organic and organic fluids are allowed.
 If using combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
 For fluids with higher viscosity ( 5 mm
speed can be adjusted.
 Fluids with viscosity over 20mm
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
®
7800 to "warm up".
TitroLine
Measuring input 1 (analog):
pH/mV-input with 24 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860).
Reference electrode 1 x 4 mm socket.
Adjustable damping settings of the pH/mV measuring signal.
RFID receiver for SI Analytics
Measurement range
pH
pH
- 3.0 ... 18.00
U mV
mV
- 2000 ... 2000
Measuring input (Pt 1000):
Temperature sensor - connector for a resistance thermometer Pt 1000
and NTC 30 kOhm. Connection: 2 x 4 mm sockets.
Measuring input (μA):
Karl Fischer (Dead-Stop) connector (µA) for double platinum electrode.
Polarisation voltage variably adjustable from 40 ... 220 mV.
Connector: 2 x 4 mm – sockets.
* The measurement uncertainty of the sensor probe has to be taken into account as well.
®
7800
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
Over range
- 3.1 ... 18.00
- 2020 ... 2020
Measurement
range T [°C]
Pt 1000
- 75 ... 195
- 40 ... 0
NTC 30
0 ... 125
Measurement
range I [A]
100
50
10
5
®
ID electrodes.
Measurement accuracy*
Display resolution
0.001
0.1
Measurement accuracy* without
Display resolution
0.1
0.1
0.1
Measurement accuracy* without
Display resolution
0.1
0.1
0.1
0.1
(Release: 16. August 2018)
without sensor probe
 1 Digit
0.002
 1 Digit
0.10
sensor probe
0.2 K  1 Digit
1.0 K  1 Digit
0.3 K  1 Digit
sensor probe
-5 /+ 3 A  1 Digit
+/- 3 A  1 Digit
+/- 1 A  1 Digit
+/- 0.2 A  1 Digit
Input
resistance 
13
> 1  10
13
> 1  10