Select the TITRONIC
500 with address «3». Confirm with <ENTER/OK> (Fig. 232).
Fig. 232
The dosing and filling rate is set as before in the dosing parameter. All other titration parameters are set as
before.
When starting the method, the external burette is automatically filled at the beginning. Then the usual queries for
sample designation and sample weight take place.
Then the titration begins:
The volume is displayed on the external device during titration. On the display of the titrator only information is
given that is dosed externally (Fig. 233).
Fig. 233
The titration curve and results are displayed as usual (Fig. 234).
Fig. 234
Of course you can also carry out a pre-titration with the external burette.