Titration parameters

4.6.5
The «Titration parameter» submenu is used to determine the actual parameters of the method (Fig. 143 and
Fig. 144).
Fig. 143
Fig. 144
4.6.5.1 Generally applicable titration parameters
Depending on the titration mode (dynamic, linear, End-Point titration, Dead-Stop titration and pH-Stat titration),
it is possible to enter a variety of parameters.
The following parameters are valid for all automatic titration modes:
Measured value (pH, mV, µA)
Measurement speed
Initial waiting time
Pre-titration
Titration end
The measurement speed and the titration end differ again as a function of the respective titration mode.
The «Measured value» is the first selection to be made, e.g. «pH (A)» (Fig. 145).
Fig. 145
215