Xylem TitroLine 7800 Operating Manual page 37

Table of Contents

Advertisement

181
Once calibration completed, the display will show the slope and the zero point of the electrode (Fig. 60).
Fig. 60
The calibration values will be automatically printed or stored as a PDF file.
<ESC> will take you back to the main menu.
The current calibration values can be viewed at any time.
Press <CAL> in the main menu. The display changes (Fig. 61).
Fig. 61
Press <MODE> (Fig. 62).
Fig. 62