Xylem TitroLine 7800 Operating Manual page 40

Table of Contents

Advertisement

184
These stages can also be changed during manual titration (Fig. 67).
Fig. 67
Stage 5 corresponds to maximum titration speed. Speed is reduced by 50% each time.
Example:
WA 20 interchangeable unit
Stage 5
100 %
Stage 5
50 %
Stage 4
25 %
Stage 3
12.5 %
Stage 2
6.8 %
Stage 1
3.4 %
Even if the titration is completed, press <STOP> or approx. for 1 sec. the grey key of the manual controller.
The titration result will be calculated and displayed (Fig. 68).
Fig. 68
The result can also be printed or stored in PDF- and CSV-format.
<ESC> will take you back to the start menu way to start the next titration immediately.
Filling of the interchangeable unit occurs automatically.
(ca.
40 ml/min)
(ca.
20 ml/min)
(ca.
10 ml/min)
(ca.
5 ml/min)
(ca. 2.5 ml/min)
(ca. 1.25 ml/min)