Download Print this page

Functions Menu Setup; Ch1, Ch2; Ch3, Ch4 - Tenma 72-7630 Instructions Manual

4 channel digital storage oscilloscope

Advertisement

Functions Menu
From
From Edge
To
To Edge
OK
Note: Customised parameters are for quick
parameter measurement. The measurement menu
of your oscilloscope is capable of measuring 26
waveform parameters. Two of them (delay and phase
for advance measurement) are generally used when
you do not require to measure all parameters but only
a few. You can set these as customized parameters.
Customized parameters are displayed on the screen.
Automatic measurement of voltage parameters
The DSO series oscilloscope can automatically measure the following voltage
parameters.
Peak to peak value (Vpp) - the voltage value from the highest point to lowest point
of the waveform.
Maximum value (Vmax) - the voltage value from the highest point to ground of the
waveform.
Minimum value (Vmin) - the voltage value from the highest point to ground of the
waveform.
Amplitude value (Vamp) - the voltage value from top to base of the waveform.
Mean (Vmid) - half of the amplitude value.
Top value (Vtop) - the voltage value from the level top to ground of the waveform.
Base value (Vbase) - the voltage value from the level base to ground of the
waveform.
Overshoot - the ratio value of the difference between maximum value and top value
of the waveform to the amplitude value.
Pre-shoot - the ratio value of the difference between minimum value and base value
of the waveform to the amplitude value.
Average value - average amplitude of signals within 1 cycle.
RMS value - the effective value. Energy generated by AC signal conversion during
one cycle with respect to the DC voltage that produces equivalent energy.
Automatic measurement of time parameters
The DSO series oscilloscope can automatically measure the frequency, cycle, rising
time, falling time, positive pulse width, negative pulse width delay positive duty ratio
negative duty ratio and nine time parameters.
Definitions are as follows:
Setup

CH1, CH2

Select one channel waveform as the delayed reference

CH3, CH4

waveform
Take the centre point of 10%-90% of the waveform rising or
Rise, Fall
falling edge
CH1, CH2
Select one channel waveform as the delayed measurement
CH3, CH4
waveform
Take the centre point of 10%-90% of the waveform rising or
Rise, Fall
falling edge
Selection of the measurement point of delayed measurement
waveform is complete and return to the measurement menu
Notes
22