How To Set Sampling Mode - Owon VDS6000 Series User Manual

2ch pc oscilloscopes
Hide thumbs Also See for VDS6000 Series:
Table of Contents

Advertisement

Vbase
voltage between flat base and ground (GND)
Overshoot
equals (Vmax - Vtop) / Vamp
Preshoot
equals (Vmin - Vbase) / Vamp
the arithmetic average voltage value of complete measured signal,
Vavg
or chosen part of measured signal
true RMS voltage value over complete measured signal, or chosen
Vrms
part of measured signal
automatic measurement options towards time measurement
10 options involved: Frequency (abbreviated as "Freq"), Period, Rise Time, Fall Time, +Width, -Width,
Delay 1→2 , Delay 1→2 , +Duty and -Duty.
Following illustration to assist better understanding different measurement options,
Definitions about these options -
the time span for signal amplitude rise from 10% to 90%, judged by the rising edge
Rise Time
of its first pulse
the time span for signal amplitude fall from 90% to 10%, judged by the falling edge
Fall Time
of its first pulse
+Width
the pulse width that the first positive pulse at 50% amplitude point
-Width
the pulse width that the first negative pulse at 50% amplitude point
Delay 1→2
the time delay of Channel 1 and Channel 2 at rising edge
Delay 1→2
the time delay of Channel 1 and Channel 2 at falling edge
+Duty
measured by +Width / Period
-Duty
measured by -Width / Period
vii. how to set sampling mode
From Function menu, via
, enter into "Sampling" option.
31