tyPE= (Remote Serial Data Type)
This setting selects the format of data received at the serial port in remote display mode.
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, press the Fn/▲ key to scroll through the available values.
When the desired value is displayed, press the TARE key to select the displayed value
and proceed to the next prompt.
Available values are:
0 = SMA - No remote key presses transmitted
1 = SB-400 - Transmits remote key presses
2 = SB-200 - Transmits remote key presses
3 = Toledo Short - No key presses transmitted
4 = Toledo Long - No key presses transmitted
5 = Rice Lake IQ355 - No key presses transmitted
After selecting the Remote Serial Data Type, the next prompt displayed will be baud=.
Proceed to the bAUd= (Serial Port Baud Rate) section to configure the baud rate, parity, and
stop bits for the remote display function.
USA= (Domestic or International)
This is the prompt to select whether the indicator is used in the USA (domestic) or outside the
USA (international).
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, press the Fn/▲ key to scroll through the available values.
When the desired value is displayed, press the TARE key to select the displayed value
and proceed to the next prompt.
Available values are:
USA= YES (Domestic)
Overcapacity status is on when
Gross weight is greater than 104%
of capacity.
Zero range is up to capacity when
trL=no.
Piece counting is disabled if
LFt=yES.
Power up zero is limited to plus or
minus 20% of capacity.
LFT= (Legal For Trade)
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, press the Fn/▲ key to scroll through the available values.
When the desired value is displayed, press the TARE key to select the displayed value
and proceed to the next prompt.
Available values are:
YES
Legal For Trade
NOTE: When both LFt=yes and USA=yes:
Piece count operation will be disabled.
NOTE: When both LFt=no and USA=yes:
Piece count operation will be enabled.
8400-0181-OM Rev C 185 and 185B Installation & Technical
16
USA= No (International)
Overcapacity status is on when
Gross weight is greater than
capacity plus 9 divisions.
Zero range is limited to plus or
minus 2% of capacity.
Piece counting is enabled
no
Not Legal For Trade