Entering Pressure Calibration Data

To enter calibration data for the pressure transducer on the TF
Probe or MISP Probe, you must list the zero and span range in
mV (TF Probe) or FP (M2SP Probe) as well as the zero and span
range in PSIg or kPa.
1. To enter the Pressure window, press the [ ] arrow key to
reach the Pressure tab and press
similar to Figure 3-4 below.
Hygro
Zero
Span
Figure 3-4: The Pressure Window in the Calibration
2. Press the [
mV) in the calibration table.
3. Press
[ENTER]
enter the desired value, and press
entry.
4. Press the [ ] key to move to the Pressure text box (here
shown as kPa), and press
keys to enter the desired value, and press
the entry.
5. Use the arrow keys to move to the Span boxes, and repeat
steps 3 and 4 to enter the Span values.
6. When you have completed entering values, press
The PM880 returns to the Pressure tab at the top of the screen.
Programming Site Data
Calibration - [0000000]
Pressure
Saturation Probe ID
mV
0.00
0.00
Option
] key to enter the first zero value (here shown as
to open the text box. Use the numeric keys to
[ENTER]
. The screen appears
[ENTER]
kPa
0.00
0.00
Exit
to confirm the
[ENTER]
. Again, use the numeric
[ENTER]
August 2003
to confirm
(Exit).
[F3]
3-9