GE PM880 Operations & Installation Manual page 160

Portable hygrometer
Hide thumbs Also See for PM880:
Table of Contents

Advertisement

August 2003
Updating Software Via IrOBEX (cont.)
5. The meter now asks if you wish to reload the flash memory
via the infrared communications port.
Press
[F1]
Press
[F2]
A series of questions ask if you wish to erase all sites, the
language pack, and the help text. Press
sure the infrared sensor on the meter has a clear path to the
infrared adapter on the PC. Now return to the PC.
6. You now have two options:
In Windows Explorer, browse to the file you wish to send.
Select and right-click on it. Select Send To
Recipient or Nearby Computer (or QuickBeam if
QuickBeam has been installed).
Open the Infrared Transfer Application (My
Computer\Infrared Recipient), click on Send Files, and
select the files you wish to send.
In either case, a window displays the status of the file upload.
8-14
(Yes) to begin reloading via OBEX, or
(No) to stop the reloading procedure.
(Yes) or
[F1]
[F2]
Infrared
Servicing the PM880
(No). Be