Specifications

7.10 Capacitance Specifications

Range and Resolution
60.00 nF, 600.0 nF, 6.000 μF
60.00 μF, 600.0 μF
2000 μF
1. Accuracy specified with film capacitor or better
2. Temperature coefficient: 0.25 x (specified accuracy) / °C [1.8°F] @ (0 ~ 18°C [32 ~ 64.4°F] or 28
~ 40°C [82.4 ~ 104°F])
3. Temperature coefficient: 0.25 x (specified accuracy) / °C [1.8°F] @ (0 ~ 18°C [32 ~ 64.4°F] or 28
~ 40°C [82.4 ~ 104°F])

7.11 Diode Specifications

Range and Resolution
1.000 V
Test Current: 0.56 mA, typical
Open Circuit Voltage: < 1.8 V DC typical

7.12 DC Current Specifications (Clamp)

Range and Resolution
200.0 A
0 ~ 500 A
500 ~ 2000 A
1. Induced error from adjacent current-carrying conductors: < 0.1 A/A
2. Accuracy specified with DC Zero (Relative) mode applied to offset non-zero residual readings

7.13 AC Current Specifications (Clamp)

Range and Resolution
200.0 A
0 ~ 500 A
500 ~ 2000 A
#NAS100019; r. AC/59959/59961; en-US
Accuracy
± (2.0% + 5 digits)
± (3.5% + 5 digits)
± (4.0% + 5 digits)
Accuracy
± (1.0 + 3 digits)
Accuracy
± (2.0% + 5 digits)
± (2.0% + 5 digits)
± (3.0% + 5 digits)
Frequency
50 ~ 60 Hz
1
2
3
1 2
Accuracy
1
± (2.0% + 5 digits)
± (2.5% + 5 digits)
± (3.0% + 5 digits)
26