5
Meter Operation
Figure 5.5 Continuity Measurements. Note the CLOSED (0.00) and OPEN (OL) circuit
examples.
1. Set the function switch to the Continuity
2. If necessary, short press the MODE button to select the Continuity
function.
3. Insert the black test lead into the negative (COM) terminal and the red test
lead into the positive terminal.
4. Place the probe ends of the test leads in parallel to the part under test.
5. If the measurement is <10 Ω, the meter will beep. If the measurement is >
200 Ω, the meter will not beep. Between 10 Ω and 200 Ω, the meter will
stop beeping at an unspecified point.
5.13 Diode Measurements
Do not perform diode tests before removing power to the diode under test. Injury to per-
sons can occur.
#NAS100019; r. AC/59959/59961; en-US
WARNING
position.
16