9.2.2

Faults

The symbol
precedes each fault.
The LED "not ready" flashes on the control panel. One of the following error messages appears
when you press "ENTER".
Number
Error message
in display
1
Slave faulty
2
Hardware fault
3
Heating has switched off
4
EEPROM jumper must be set
8
Parameter storage test not carried out
9
EPROM faulty
19
Sensitivity too low
40
Oven temperature outside limits or Pt100
faulty
41
Pump temperature outside limits or Pt100
faulty
42
Sample gas probe temperature outside
limits or Pt100 faulty
43
Sample gas line temperature outside limits
or Pt100 faulty
44
Flame monitor faulty
45
Flow sensor temperature outside limits or
Pt100 faulty
46
Flow outside limits or sensor faulty
47
Sample gas pressure too high or too low or
pressure sensor faulty
Comment
Pull mains plug immediately,
inform service
Inform service
RESET
Inform service
u
See
8, next page but one
Order replacement from
manufacturer
Incorrect calibration gas,
incorrect measuring range,
see functions 2 and 6
Check contacts
Check contacts
Check contacts
Check contacts
Check contacts
Check contacts
Check contacts
Check cylinder pressure,
readjust, check pressure sensor