Siemens FIDAMAT 5E Instruction Manual page 55

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

Autocalibration
C
** 47 Set sample gas time between zero and
sensitivity calibrations;
set calibrations to be omitted
CL
1
4
7
Identifier
Set calibrations
to be omitted
2
Identifier
3
Identifier
** 48 Signal contact at start/end of "autocal"
CL
Sample gas time,
max. 99 min
Zero calibrations to be
omitted, max. 99
Sensitivity calibrations
to be omitted, max. 99
4
8
1
Start
0
End
A-19
This function is selected if zero and sensitivity
calibrations are not to be carried out directly in
succession. The total time between the zero and
sensitivity calibrations comprises the sample
gas time and the time selected under function
44, identifier 2.
Factory setting: 0
If a number is entered under identifier 2 for zero
calibrations to be omitted, the number under
identifier 3 is automatically set to zero and vice
versa.
Factory setting: 0/0
Plug connector X2, pins 9 and 10
Factory setting: 1
C