Repeatability
Characteristic
Long-term drift
Zero
5E-E
5E-A, 5E-AS
5E-I
Span
Response time (T
time) with time
90
constant 0 s, set using function 13
Warm-up time
Analyzer chamber temperature
Permissible storage temperature
Permissible ambient temperature (operation)
Inlet pressure for hydrogen, combustion air,
zero gas and calibration gas(es)
Position of use
Change in display through influencing variables
Mains voltage variations (+10, -15 %)
Ambient temperature variations
Atmospheric pressure variations
Frequency variations
Influence of position
Power supply
Mains connection
Power consumption
Degree of protection
Rack-mounted unit
Bench-top unit
Dimensions
Weight
Material of parts in contact with sample gas
0.1 to 1 % of full-scale value, range-dependent
Linearization error <1 % of f.s.v.
<2 % of span/week with f.s.v. >10 ppm C
range-dependent
<2 % of span/week with f.s.v. >10 ppm C
range-dependent
< 2 % of span/day with f.s.v. 2 ppm C
dependent
As for zero
<1 s without filter
2 to 3 s with filter
Approx. 2 to 3 hours
110 to 200 ˚C, freely-selectable
–30 to +70 ˚C
5 to 45 ˚C
3 bar (45 psig) above atmospheric pressure
Front panel vertical
<1 % of measured value
<1 % of measured value/10 K
<1 % of measured value /50 mbar
<1 % of measured value for the permissible
frequency range
<1 % of measured value with inclination <15˚
AC 110, 120, 220, 230, 240 V, 48 to 63 Hz
Approx. 350 VA in start-up phase,
approx. 150 VA during operation
IP 20 to DIN 40050
IP 21 to DIN 40050
See dimensional drawing
28 kg
PTFE, quartz (nozzle, fused silica), graphite
(gaskets), stainless steel 1.4571 (316 SS)
9
,
1
,
3
, range-
1