Siemens FIDAMAT 5E Instruction Manual

Gas analyzer for measurement of total hydrocarbon content

Advertisement

FIDAMAT 5E gas analyzer
for Measurement of Total Hydrocarbon Content
7MB1420
Instruction Manual
Order no. C79000-B5276-C106-04

Advertisement

Table of Contents
loading

Summary of Contents for Siemens FIDAMAT 5E

  • Page 1 FIDAMAT 5E gas analyzer for Measurement of Total Hydrocarbon Content 7MB1420 Instruction Manual Order no. C79000-B5276-C106-04...
  • Page 2: Table Of Contents

    Mode of Operation Technical Data Installation Mounting Gas Supply Electric Connections Startup, Calibration Startup 7.1.1 FIDAMAT 5E-E, 5E-A 7.1.2 FIDAMAT 5E-I 7.1.3 FIDAMAT 5E-AS Calibration Automatic Calibration Operation Description of Control Panel Input Example Coding and Organization of Operating Function Groups...
  • Page 3 All commitments on the part of Siemens are contained in the respective sales contract which also contains the complete and solely applicable warranty conditions. These warranty conditions in the contract are neither extended nor limited by the contents of this Instruction Manual.
  • Page 4: Introduction

    Introduction General Warning! Certain parts in this electrical device contain dangerous voltages during operation. Failure to observe the warnings could therefore result in severe personal injury and/or damage to property. Only suitably qualified staff should work on this device or in its vicinity. These should be thoroughly familiar with all warnings and maintenance measures according to this Manual.
  • Page 5: Application

    Application The FIDAMAT 5E is an analyzer for measuring total hydrocarbons. It operates according to the principle of flame ionization. The analyzer is suitable for measuring all gaseous hydrocarbons. A heated line must be used for the connection to prevent condensation and adsorption in the gas line.
  • Page 6: Design

    Design The device is available either as a bench-top unit fitted in a housing, or without a housing for installation in a 19-inch rack or 19-inch cabinet. The FIDAMAT comprises two main sections, the analyzer section and the electronics with front panel controls.
  • Page 7: Mode Of Operation

    Mode of Operation Electrons are released when hydrocarbons are combusted in a hydrogen flame. These electrons are collected at an electrode by means of an electric field and measured using a highly sensitive amplifier. The current is proportional to the quantity of organically-bound C atoms in the sample gas.
  • Page 8: Technical Data

    Technical Data Number of measuring ranges Max. 4, freely parameterizable Switching ratio Optional Autoranging Selectable Hysteresis: – Switchover to larger range at 90 % of span of smaller range – Switchover to smaller range at 80 % of span of smaller range Output signals At terminal strips X1, X2 and X3.
  • Page 9 Repeatability 0.1 to 1 % of full-scale value, range-dependent Characteristic Linearization error <1 % of f.s.v. Long-term drift Zero 5E-E <2 % of span/week with f.s.v. >10 ppm C range-dependent 5E-A, 5E-AS <2 % of span/week with f.s.v. >10 ppm C range-dependent 5E-I <...
  • Page 10 Gas consumption data 5E-E Combustion gas Approx. 350 ml/min (approx. 21 l/h) Hydrogen Approx. 20 ml/min (approx. 1.2 l/h) Sample gas Approx. 1 l/min (approx. 60 l/h) Zero gas/calibration gas Approx. 2 l/min (approx. 120 l/h) 5E-A (sample gas at standard pressure) Combustion gas Approx.
  • Page 11 FIDAMAT 5E rack-mounted unit Caution: Only mount FIDAMAT 5E on sliding rails or angled rails Sample gas connection: 6 mm screw connection or 8 mm screw connection or 1/4" screw connection Sample gas outlet: 8 mm PVDF screw connection Auxiliary gas connections: 6 mm screw connection or 1/4"...
  • Page 12: Installation

    FIDAMAT 5E-E and 5E-I and 600 mbar abs. for the FIDAMAT 5E-A. The exhaust line from the FIDAMAT 5E must always be routed with a downward gradient since water condenses in it. The internal diameter should not be less than 8 mm (the connection is provided for a hose 8 x 1 (10 mm external diameter)).
  • Page 13: Electric Connections

    3 Calibr. gas 4 Zero gas Calibr. gas 1 Fig. 6.1 Rear panel of FIDAMAT 5E Electric Connections Warning! The following must be observed for the electric connections: VDE 0100 “Regulations for the installation of power systems with mains voltages below 1000 V”.
  • Page 14 Pin assignment diagram FIDAMAT 5E-E Device side Connection side Device side Connection side MR 1 Analog current + MR 2 Analog current – Measuring range identification MR 3 Vacant Contact closed = MR active MR 4 Analog current – Calibrate zero (funct. 5) MR 4 Calibrate span (funct.
  • Page 15 Pin assignment diagram FIDAMAT 5E-I Device side Connection side Device side Connection side MR 1 Analog current + MR 2 Analog current – Measuring range identification MR 3 Vacant Contact closed = MR active MR 4 Analog current – Calibrate zero (funct. 5) MR 4 Calibrate span (funct.
  • Page 16 Pin assignment diagram FIDAMAT 5E-A Device side Connection side Device side Connection side MR 1 Analog current + MR 2 Analog current – Measuring range identification MR 3 Vacant Contact closed = MR active MR 4 Analog current – Calibrate zero (funct. 5) MR 4 Calibrate span (funct.
  • Page 17 Pin assignment diagram FIDAMAT 5E-AS Device side Connection side Device side Connection side MR 1 Analog current + MR 2 Analog current – Measuring range identification MR 3 Vacant Contact closed = MR active MR 4 Analog current – Calibrate zero (funct. 5) MR 4 Calibrate span (funct.
  • Page 18 Pin assignment diagram Pt100 for heater of sample gas line Pt100 for heater of sample gas line Control output for heater of sample gas line +24 V Control output for heater of sample gas line Vacant Pt100 for heater of sampling probe Pt100 for heater of sampling probe Control output for heater of sampling probe +24 V Control output for heater of sampling probe...
  • Page 19: Startup, Calibration

    1000 mbar. The pressures need not be reset following a restart. 7.1.2 FIDAMAT 5E-I With the FIDAMAT 5E-I, the sample gas pressure (approx. 550 mbar) is set automatically. The hydrogen pressure is set or corrected in the same manner as with the FIDAMAT 5E-E and 5E-A.
  • Page 20: Fidamat 5E-As

    7.1.3 FIDAMAT 5E-AS With the FIDAMAT 5E-AS, the excess sample air is output from the analyzer via the bypass. In calibration mode, the excess calibration gas is output via the gas outlet. Only one zero gas and one calibration gas are available for the analyzer.
  • Page 21 Sample Set time up to first automatic mode calibration (function 41) Start Zero gas valve opens, signal contact “Start calibration” Zero gas purging Purging time Function 44, identifier 1 Zero point being calibrated Calibration Zero gas valve closes; LED “autocal” flashes Calibration gas valve Calibration LED “not ready”...
  • Page 22 Function 44 must be parameterized for the slave. The slave can trigger the automatic calibration procedure for a further analyzer at the end of its calibration etc. Connections Slave Master An automatic calibration can be triggered at any time using function 40. Function 22 must not be activated in this case.
  • Page 23: Operation

    ENTER Limit No. Fig. 8.1 Control panel of FIDAMAT 5E Display of selected measuring range 1 to 4 Large display: concentration display, parameter display, fault display, status display Small display: display of selected function codes; if the small display is dark, the large...
  • Page 24: Input Example

    Function table for the most important functions Switchover from measurement to calibration mode and simultaneous signalling of these statuses via a contact Operator prompting: the corresponding LED flashes here if further inputs are required for the entered functions Every input must be terminated by the “ENTER” key apart from the code number (10) Flame monitor: The LED lights up when the flame is burning and goes out when the...
  • Page 25: Coding And Organization Of Operating Function Groups

    Coding and Organization of Operating Function Groups The functions are combined into groups according to their type: Function groups Types of function Display functions Calibration and adjustment functions Functions for input of application parameters Functions for input of device parameters To prevent important parameters from being deleted or modified by mistake, there are several function levels which can only be accessed using special codes.
  • Page 26: Organization Of Data Storage

    User-specific data Current data. All parameters required to operate the FIDAMAT 5E are stored in the EPROM and are protected by hardware (basic factory-set data). These data can be loaded into the current parameter memory using function 53. The user-specific data set is overwritten in the process, however. The basic factory-set data can only be overwritten by a hardware intervention.
  • Page 27: Maintenance, Warnings And Error Messages

    Maintenance, Warnings and Error Messages Maintenance In the FIDAMAT 5E it is only necessary to regularly service the pump. It is recommendable to replace the pump diaphram every six months. The pump can be removed from the front of the analyzer.
  • Page 28: Warnings And Error Messages

    Warnings and Error Messages 9.2.1 Warnings The symbol precedes each warning. The LED “warning” flashes on the control panel. The cause of the warning can be scanned using “ENTER”. Further measurements are possible in this state. The analyzer may fail if the cause of the warning is not eliminated.
  • Page 29 Number Cause of warning Causes, notes in display Sample gas pressure outside the limits The limits defined in function 38 have been violated. Combustion air pressure (con- trol air) is too high or too low; sample gas inlet faulty (pump, filter, lines);...
  • Page 30: Faults

    9.2.2 Faults The symbol precedes each fault. The LED “not ready” flashes on the control panel. One of the following error messages appears when you press “ENTER”. Number Error message Comment in display Slave faulty Pull mains plug immediately, inform service Hardware fault Inform service Heating has switched off...
  • Page 31 Number Error message Comment in display Hydrogen pressure too high or too low or Check cylinder pressure, pressure sensor faulty readjust, check pressure sensor High-voltage outside limits Check whether high-voltage line is connected Operating voltages outside setpoints Check power supply and con- nections on transformer Flow too high Check outlet restrictor,...
  • Page 32 Number Cause of error Comment in display Parameter storage test not carried out A: Analyzer does not react to ENTER Press dot key for 30 sec until function 50 appears in the small display; recode the analyzer. Then reload basic data into RAM using function 53 or 55.
  • Page 33: Spare Parts And Accessories

    Flame ionization detector, complete, 5E-E and 5E-I C79451-A3405-B510 Flame ionization detector, complete, 5E-A and 5E-AS C79451-A3405-B578 FID cover complete with thermocouple line C79451-A3405-B501 Quartz nozzle for FIDAMAT 5E-E and 5E-I C79402-Z1282-C1 Quartz nozzle for FIDAMAT 5E-A and 5E-AS C79402-Z1282-C2 Clamping screw in the FID C79211-A3003-C15...
  • Page 34 Order No. Crimp restrictor 2, combustion air C79451-A3405-C153 Crimp restrictor 3, auxiliary air, 5E-E C79451-A3405-C154 Crimp restrictor 3, auxiliary air, 5E-A C79451-A3405-C164 Crimp restrictor 3, auxiliary air, 5E-AS C79451-A3405-C224 Crimp restrictor 4, exhaust, 5E-E and 5E-AS C79451-A3405-C155 Crimp restrictor 4, exhaust, 5E-A C79451-A3405-C168 Crimp restrictor 5, zero gas, calibration gas(es) C79451-A3405-C156...
  • Page 35 Order No. Solenoid valve for for H , combustrion air, control air, bypass gas C79402-Z871-A5 Solenoid valve for for H , zero gas, calibration gases C79402-Z871-A7 Gasket, diameter 4 mm C79451-A3040-C331 Gasket, diameter 6 mm C79451-A3040-C354 Gasket, diameter 10 mm C79451-A3040-C353 O-ring, diameter 10 mm F79402-E871-A1...
  • Page 36 Order No. Display board (front panel) C79451-A3210-A502 Touch pad keyboard with labels, German C79451-A3405-D52 Touch pad keyboard with labels, English C79451-A3405-D53 Touch pad keyboard with labels, French C79451-A3405-D54 Touch pad keyboard with labels, Spanish C79451-A3405-D55 Touch pad keyboard with labels, Italian C79451-A3405-D56 See Catalog PA 10 for accessories Heated sample gas line...
  • Page 37: Appendix

    Solenoid valves 1 and 6 - 10 closed in de-energized state Zero Calibration gas 1 Calibration gas 2 Calibration gas 3 Calibration gas 4 optional Fig. 11.1 FIDAMAT 5E-E for hydrocarbon measurements in the chemical industry, automobile industry and emission plants...
  • Page 38 Pressure sensor Restrictors Damping Pump Filter Sample Pump 110 to 190 ˚C Oven 110 to 200 ˚C All solenoid valves closed in de-energized state Zero Calibration gas 1 Fig. 11.2 FIDAMAT 5E-I for ambient measurements of hydrocarbons (e.g. TLV measurements)
  • Page 39 Solenoid valve 2 open in de-energized state Solenoid valves 1 and 6 - 10 closed in de-energized state Zero gas Calibration gas (1) Calibration gas 2 Calibration gas 3 Calibration gas 4 optional Fig. 11.3 FIDAMAT 5E-A for measurements using H...
  • Page 40 Excess flow during calibration Oven (heated up to 230 ˚C) Solenoid valves 2 and 3 open in de-energized state all other solenoid valves closed in de-energized state Zero gas Calibration gas (1) Option Fig. 11.4 FIDAMAT 5E-AS for shed measurements...
  • Page 41: Layout Diagrams Of Important Electronic Components

    11.2 Layout Diagrams of Important Electronic Components Master Slave Motherboard Cooling rail EEPROM protection switch Terminal strips Control board Heat sink Potentiometer for adjustment of flow sensor Fig. 11.5 Motherboard (top) and control board...
  • Page 42: Summary Of Operating Functions

    11.3 Summary of Operating Functions Symbols in the sequence diagrams Display of set value and possible adjustment for a new value. When entering values make sure to enter the sign first! ENTER key (importing of number) Display output in large display (the measured value is displayed again after approx.
  • Page 43 Survey of operating functions Display and switch over measuring ranges, switch over display 1 Change measuring range 2 Display measuring range ** 20 Autoranging on/off * 94 Switch over between mg/m and ppm Calibration of measuring ranges 5 Calibrate zero 6 Display sensitivity setpoint 7 Adjust sensitivity setpoint 8 Calibrate senstitivity to setpoint...
  • Page 44 Autocalibration (continued) 46 Display time up to next “autocal” ** 47 Set sample gas time between zero and sensivity calibrations; set calibrations to be omitted ** 48 Signal contact at start/end of “autocal” Temperatures, pressures, flow, limits 9 Display limits ** 18 Set limits to upward or downward exceeding ** 19 Set limit ** 23 Limit monitoring on/off...
  • Page 45 Time constants ** 13 Set time constant T ** 14 Set dynamic time constant T (dynamic noise rejection) Measured-value memory, current output ** 21 Measured-value memory on/off ** 24 Set current output Redefine measuring ranges ** 11 Define start-of-scale values of measuring ranges ** 12 Define full-scale values of measuring ranges ** 51 Select number of valid measuring ranges Further display functions...
  • Page 46 Test functions, RESET, switch-off (continued) * 29 Measure during warming-up phase ** 49 Switch valves individually *** 53 Load basic factory-set data *** 55 Load user-specific data *** 57 Store user-specific data *** 62 Reproduce internal values on analog signal, channel 1 *** 65 STANDBY *** 67 Ignite flame *** 68 RESET...
  • Page 47 1 Change measuring range 2 Display measuring range 5 Calibrate zero 6 Display sensitivity setpoint 7 Adjust sensitivity setpoint 8 Calibrate senstitivity to setpoint 9 Display limits ** 11 Define start-of-scale values of measuring ranges ** 12 Define full-scale values of measuring ranges ** 13 Set time constant T ** 14 Set dynamic time constant T (dynamic noise rejection)
  • Page 48 ** 34 Display, set max. temperature deviation * 35 Display, set H pressure * 36 Display, set sample gas pressure * 37 Enter tolerance for hydrogen pressure * 38 Enter tolerance for sample gas pressure ** 39 Probe and/or heated line present/not present ** 40 Trigger single “autocal”...
  • Page 49 *** 67 Ignite flame *** 68 RESET 69 LED check, software No. *** 71 Compensation of temperature influence on zero *** 73 Compensation of temperature influence on measured value *** 77 Suppression of brief noise signals *** 78 Suppression of negative measured values *** 79 Pause * 92 Display warnings and brief errors * 94 Switch over between mg/m...
  • Page 50 Display and switch over measuring ranges, switch over display There are four measuring ranges. The measu- 1 Change measuring range ring range which is switched on is indicated to the right of the display. The number of active measuring ranges can be defined using function The start-of-scale and full-scale values of the measuring range can be freely selected.
  • Page 51 Calibration of measuring ranges Zero gas is applied after pressing 5 and ENTER 5 Calibrate zero until the displayed measured value is imported as the new zero by pressing ENTER again. The LED “not ready” lights up during the adjustment. Display The zero of all measuring ranges is calibrated.
  • Page 52 Calibration of measuring ranges Total calibration: ** 17 Switch over from total calibration to individual The calibration applies to all four measuring calibration ranges. Individual calibration: Each measuring range is calibrated separately. If four calibration gas connections are present on the analyzer, note that range 1 is calibrated with Individual calibration calibration gas 1, range 2 with calibration gas 2...
  • Page 53 Autocalibration Autocal has been programmed using functions ** 22 "autocal" on/off 40 ff. The LED “autocal” flashes during the automatic calibration. The calibration is repeated at an interval defined in function 42. On (activated) Factory setting: 0 (not activated) An automatic calibration (zero, calib ration ** 40 Trigger single “autocal”...
  • Page 54 Autocalibration The flow of calibration gas is enabled during the ** 44 Set purging times purging time. The time parameterized under identifier 2 is also used as the prepurging time for the sample gas following the calibration. Identifier: n zero gas Identifier Purging time calibration gas (1)
  • Page 55 Autocalibration This function is selected if zero and sensitivity ** 47 Set sample gas time between zero and calibrations are not to be carried out directly in sensitivity calibrations; succession. The total time between the zero and set calibrations to be omitted sensitivity calibrations comprises the sample gas time and the time selected under function 44, identifier 2.
  • Page 56 Temperatures, pressures, flow, limits This function displays the limit associated with 9 Display limits the limit contact. The measuring ranges are dis- played (LEDs) in which the limit is active. Upward violation Display Downward violation Limit contact No., the associated symbol for upward/downward violation appears to the left of the number The selected limit contact can be set to upward...
  • Page 57 Temperatures, pressures, flow, limits The temperatures of the oven are in ˚C. The ** 30 Display, enter actual temperature and setpoint ** 30 Display, enter actual temperature and setpoint minimum value is 110 ˚C, the maximum value temperature of oven temperature of oven 200 ˚C.
  • Page 58 Temperatures, pressures, flow, limits This function defines how far the temperatures ** 34 Display, set max. temperature deviation of the oven, pump, heated sample gas line and sample gas probe may deviate upwards or downwards from the setpoint temperature. A warning is output if the temperature deviates by more than this value for one of the four heaters.
  • Page 59 Temperatures, pressures, flow, limits Identifier 1 = probe *** 39 Probe and/or heated sample gas line present/ Identifier 2 = heated line not present The device configuration is defined by this function. Factory setting: Identifier 1: 0 Identifier Not present Identifier 2: 0 Present A-23...
  • Page 60 Time constants The time constant T (0.1 to 100 s) is used to ** 13 Set time constant T reduce the noise. The signal delay increases with increasing T Factory setting: 0.0 0.0 to 100 seconds The time constant T applies to small changes ** 14 Set dynamic time constant T (dynamic noise...
  • Page 61 Measured-value memory, current output If the measured-value memory is switched on, ** 21 Measured-value memory on/off the last measured value is stored when the CAL key is pressed. It is then present at the mA output until the analyzer is reset to the measurement status by pressing the MEAS key.
  • Page 62 Redefine measuring ranges The start-of-scale values can be set as desired. ** 11 Define start-of-scale values of measuring The only condition is that the full-scale value set ranges using function 12 is greater than the start-of-sca- le value set here. The unit displayed when the setting is made applies as the unit to the value entered here (switch over unit using function 94).
  • Page 63 Further display functions The current compensated in the zero is display- 15 Display compensated background current ed here in pA. (zero calibration) Display * 60 Display internal values Identifier: n = Oven temperature (100 to 200 ˚C) Physical section temperature (10 to 60 ˚C) Pump temperature (110 to 190 ˚C) Pump temperature comparison...
  • Page 64 Adjustment and calibration functions Iden- Setpoint *** 71 Compensation of temperature influence on tifier zero Temperature of analyzer section with average ambient temperature (in ˚C; see function 60, identifier 2) Change in display with increase in temperature (function 61, identifier 4) Identifier value Change in display with reduction in...
  • Page 65 Adjustment and calibration functions Iden- Setpoint *** 73 Compensation of temperature influence on tifier measured value Temperature of analyzer section with average ambient temperature (in ˚C; see function 60, identifier 2) Change in display with increase in temperature Identifier value Change in display with reduction in temperature Procedure...
  • Page 66 Calibration gas 3 valve Calibration gas 4 valve Bypass gas valve With the FIDAMAT 5E-E and 5E-A the lines are flushed in the event of a power failure (combus- tion air valve open in de-energized state). Programming with original data (status when *** 53 Load basic factory-set data delivered from factory).
  • Page 67 Test functions, RESET, switch-off Test functions, RESET, switch-off Identifier: n = *** 62 Reproduce internal values on analog signal, channel 1 Measured value Oven temperature ˚C Physical section temperature ˚C Pump temperature ˚C Display Pump temperature comparison ˚C Identifier Sampling temperature ˚C Sample gas probe temperature ˚C Flame temperature ˚C Temperature of flowmeter ˚C...
  • Page 68 Test functions, RESET, switch-off All LEDs light up. After 5 seconds, the two soft- *** 69 LED check, software No. ware versions appear in succession in the display. The entered time corresponds to a dead time *** 77 Suppression of brief noise signals which will be effective only if the measured value exceeds the value entered under function 14, identifier 2.
  • Page 69 Test functions, RESET, switch-off * 96 Display ∆ zero calibration This function can be used to automatically com- pare the previous and new zero and sensitivity calibrations. A warning is output if the difference between the two calibrations is >6 %. The valu- es can be viewed using identifiers 3 and 4.
  • Page 70 Reprogram, activate and deactivate codes Identifiers *** 50 Reprogram, activate/deactivate codes codes are active codes are not active; all operating functions are freely accessible. The codes are automatically reactivated if Identifier a function is not selected within 15 minutes. Identifier codes are permanently inactive.
  • Page 71: Special Sheet For Storage Of Current Coding

    11.4 Special Sheet for Storage of Current Coding FIDAMAT 5E gas analyzer Function and code Remarks *** 50 Reprogram of codes Small display Code 1 Code 2 Code 3 Press for approx. 30 s Function 50 is only accessible via...
  • Page 72 A-36...
  • Page 73 A-37...
  • Page 74 Siemens AG Automation Division Process Analytics Subject to technical modifications Progress AUT V351 D-76181 Karlsruhe, Germany © Copyright Siemens AG 1993 in Automation. Order No.: C79000-B5276-C106-04 Siemens Siemens Aktiengesellschaft Printed in Germany AG 0395 0.05 Sr 72 En...

This manual is also suitable for:

7mb1420

Table of Contents