Electric Connections - Siemens FIDAMAT 5E Instruction Manual

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

The supply gases (hydrogen, combustion air, zero gas and calibration gas(es)) must be available
with an inlet pressure of 3 bar above atmospheric. A higher zero gas and calibration gas pressure
leads to an increased consumption, a lower pressure may lead to an incorrect measurement since
additional sample gas may be drawn in.
Fig. 6.1 Rear panel of FIDAMAT 5E
6.3

Electric Connections

Warning!
The following must be observed for the electric connections:
VDE 0100 "Regulations for the installation of power systems with mains
voltages below 1000 V".
A mains disconnection device must be provided in the building installation
(see rating plate for loading capacity).
Check that the local power supply agrees with that specified on the rating plate on the device.
The required signal cables must be connected to the 25-pin trapezoidal plugs X1, X2 and X3
and the 5-pin round plugs X4, X5 and X6 according to the pin assignment diagram.
Calibration
H 2 /He
gas 2
Calibration
Combustion
gas 3
air
X4
Calibr. gas 4
Zero gas
Calibr. gas 1
13
X1
Mains
X2
X3
X5
X6

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7mb1420

Table of Contents