Siemens FIDAMAT 5E Instruction Manual page 53

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

Autocalibration
C
** 22 "autocal" on/off
** 40 Trigger single "autocal"
** 41 Set time up to first "autocal"
CL
4
1
Identifier
Identifier
** 42 Set time between two "autocals"
CL
4
2
43 Display time between two "autocals"
CL
2
1
2
On (activated)
0
Off
(not activated)
4
0
1
Hours (max. 1000)
2
Minutes (max. 59)
Set time 1 to 999 h
Display
4
3
A-17
Autocal has been programmed using functions
40 ff.
The LED "autocal" flashes during the automatic
calibration.
The calibration is repeated at an interval defined
in function 42.
Factory setting: 0
An automatic calibration (zero, calib ration
gas(es)) is triggered. The autocalibration cycle
is not influenced by this.
The time is defined when the first calibration
procedure is to be automatically triggered. Furt-
her automatic calibration procedures are carried
out according to the time defined in function 42.
The calibration procedure can only be triggered
when the analyzer is in the "Measure" status.
Factory setting: 24/0
The cyclic repetition of the calibration procedure
is defined here. If a time setting which has alrea-
dy been imported is modified, the new time only
becomes effective following the next autocal.
This time can be defined using function 41.
Factory setting: 168
The time is output on the display in hours.