Siemens FIDAMAT 5E Instruction Manual page 50

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

Display and switch over measuring ranges, switch over display
A
*
1 Change measuring range
2 Display measuring range
CL
2
Measuring range No.
** 20 Autoranging on/off
CL
* 94 Switch over between mg/m
CL
9
CL
1
Measuring range No.
Display
2
0
1
Autoranging
0
Manual switchover
3
and ppm
4
0
3
mg C/m
1
ppm C
1
2
ppm C
3
3
ppm C
6
A-14
There are four measuring ranges. The measu-
ring range which is switched on is indicated to
the right of the display. The number of active
measuring ranges can be defined using function
51.
The start-of-scale and full-scale values of the
measuring range can be freely selected.
Factory setting: 4
The start-of-scale and full-scale values of the
measuring range are output in succession in the
display.
This function only applies to the recorder output
(see plug connector X1). The current measured
value is always output on the display even if it
exceeds the current measuring range.
The displayed value flashes when the measuring
range(s) is exceeded.
Factory setting: 1
The LED on the right of the display indicates the
concentration unit of the display.
Factory setting: 0