Siemens FIDAMAT 5E Instruction Manual page 57

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

Temperatures, pressures, flow, limits
D
** 30 Display, enter actual temperature and setpoint
** 30 Display, enter actual temperature and setpoint
temperature of oven
temperature of oven
CL
CL
3
3
0
0
** 31 Display, enter actual temperature and setpoint
temperature of pump
CL
3
1
** 32 Display, enter actual temperature and setpoint
temperature of heated line
CL
3
2
** 33 Display, enter actual temperature and setpoint
temperature of sample gas probe
CL
3
3
Display
actual
temp.
Enter setpoint
Enter setpoint
temperature
temperature
Display
actual
temp.
Enter setpoint
temperature
Display
actual
temp.
Enter setpoint
temperature
Display
actual
temp.
Enter setpoint
temperature
A-21
The temperatures of the oven are in ˚C. The
minimum value is 110 ˚C, the maximum value
200 ˚C.
The actual temperature is initially displayed for
approx. 3 s, and then the setpoint temperature.
Factory setting: 200
The temperatures of the pump are in ˚C. The
minimum value is 110 ˚C, the maximum value
190 ˚C.
The actual temperature is initially displayed for
approx. 3 s, and then the setpoint temperature.
Factory setting: 190
The temperatures of the heated line are in ˚C.
The minimum value is 110 ˚C, the maximum
value 200 ˚C.
The actual temperature is initially displayed for
approx. 3 s, and then the setpoint temperature.
Factory setting: 190
The temperatures of the sample gas probe are
in ˚C. The minimum value is 110 ˚C, the maxi-
mum value 200 ˚C.
The actual temperature is initially displayed for
approx. 3 s, and then the setpoint temperature.
Factory setting: 190