Siemens FIDAMAT 5E Instruction Manual page 31

Gas analyzer for measurement of total hydrocarbon content
Table of Contents

Advertisement

Number
Error message
in display
48
Hydrogen pressure too high or too low or
pressure sensor faulty
49
High-voltage outside limits
50
Operating voltages outside setpoints
51
Flow too high
52
Flow too low
53
Oven temperature setpoint exceeded or
fallen below
54
Pump temperature setpoint exceeded or
fallen below
55
Heated line temperature setpoint exceeded
or fallen below
56
Sample gas probe temperature setpoint
exceeded or fallen below
57
Combustion air/sample gas pressure too low
60
Flame does not ignite (30 min)
61
Pump does not start
Comment
Check cylinder pressure,
readjust, check pressure sensor
Check whether high-voltage line
is connected
Check power supply and con-
nections on transformer
Check outlet restrictor,
examine for leaks
Filter contaminated,
pump faulty,
line blocked
Combustion air line interrupted
Check gas flows
Check sample gas pressure
(380 to 390 mbar without pump)