9
Maintenance, Warnings and Error Messages
9.1
Maintenance
In the FIDAMAT 5E it is only necessary to regularly service the pump. It is recommendable to replace
the pump diaphram every six months. The pump can be removed from the front of the analyzer.
Ensure that the gas connections on the pump are not damaged.
A pump fault is indicated by a reduced flow (often indicated in advance by warning 66) and by
increased noise on the measured signal.
The filter cartridge must be replaced depending on the application. The gaskets and the pressure
ring must also be replaced to ensure that the filter does not leak.
27