VECTORVISION SKY NAVIGATION STATION
6.3.3
Eaton UPS
Illustration
How to Turn On the
UPS
Step
Press Power On
NOTE: Information on UPS signals is provided on page 89.
How to Turn Off the
UPS
Steps
The UPS can maintain critical functions for up to three minutes by switching to local battery power.
UPS Running Time
Battery Self-Test
Steps
Changing Batteries
Technical User Guide Rev. 2.7 VectorVision & Kolibri
s
on the front panel.
1.
Press Power Off
a
into standby mode.
2.
Unplug or remove utility power from the UPS. The UPS shuts down in five seconds.
1.
Press the Self-test button and hold it for approximately three seconds.
• Flashing battery indicator light means that the batteries need to be re-
placed.
2.
• If the battery indicator light turns off, the batteries are functioning properly.
Have the battery replaced if it has not been replaced for more than three years.
Check the battery at least once a year using the UPS battery self-test. Have the battery
replaced if the self-test fails or if the UPS reports a defective battery. Contact Brainlab
support for assistance.
a s
Figure 46
and hold the button for approximately three seconds. The UPS goes
99