KOLIBRI NAVIGATION STATION
4.7.2
UPS Signals
During a power failure, the UPS emits specific signals to indicate that the system is operating from
Background
a limited internal source of power.
Explanation of
Signals
Cause
Power failure
Shutdown performed
UPS encounters severe
overload
UPS has been turned off
Mains power has been
reactivated
LED indicators on the UPS power switch, as well as UPS audio signals, indicate the status of the
System Status
system.
The UPS code status is attached to the Kolibri Cart. Further details on the code status, and infor-
mation on UPS audio signals and LED indicators, are provided on page 87.
Technical User Guide Rev. 2.7 VectorVision & Kolibri
Signal
UPS emits a beep every
minute
UPS returns to periodic
beep
UPS emits a loud tone
Stops
No signal
Action/Comment
Signals indicate that the system is
operating from a limited internal
source of power
• Turn off UPS or reactivate
mains power
• All signals stop
UPS shuts down
All signals are reset
73