Electrical Safety Test - Medical Electrical System
Diagram Table
Area
1
2
A
B
C
D
E
F
G
H
J
K
L
1
The air clearance and creepage distances of the backlight PCB are given by the design of the PCB.
2
Functional insulation
162
Insulation Type
Basic
Reinforced or Double
2
Basic
Double
Basic
Required Distance (mm)
Reference
Voltage
Clearance
250Vac
1.6
250Vac
2.5
250Vac
2.5
250Vac
5.0
250Vac
1.6
30Vdc
2
30Vdc
2
30Vdc
1
30Vdc
1
30Vdc
1
1
900Vdc
6.5
1
900Vdc
6.5
Technical User Guide Rev. 2.7 VectorVision & Kolibri
Test Voltage
Creepage
3
1500
4.0
1500
4.0
1500
8.0
4000
3.0
1500
4
500
4
500
2
500
2
500
2
500
2800V
1
12
(2U+1000V)
2800V
1
12
(2U+1000V)