Reference Manual
00809-0100-4024, Rev CA
2.7

Mounting the process seal antenna

Figure 2-22. Process Seal Antenna Dimensions
Dimensions are in inches (millimeters).
2.7.1

Preparations

It is important that the tank flange surface is flat. The maximum deviation must be within the following
specifications as illustrated:
Flange surface tolerance:
d< ±0.05 mm
Mechanical Installation
7.87 (200)
6.30 (160) 4 in. process seal
8.58 (218) 6 in. process seal
d
21.65 (550) 4 in. process seal
25.59 (650) 6 in. process seal
Mechanical Installation
March 2020
33