Europe - Emerson Rosemount 5600 Series Reference Manual

Level transmitter non-contacting radar with hart protocol
Hide thumbs Also See for Rosemount 5600 Series:
Table of Contents

Advertisement

Product Certifications
March 2020
U
o
Intrinsically Safe
Display Output
7.84 V
Entity
parameters
Up to two
Intrinsically Safe
23.1 V
Outputs Entity
parameters
U
i
Up to two
Intrinsically Safe
30 V
Inputs Entity
parameters
Special conditions of safe use
Caution statement (on the cover of the "Ex d" compartment): "To
prevent ignition of Hazardous Atmospheres, do not open when
energized or when explosive gas is present".
Caution statement (on the cover of each "Ex e" compartment): "To
prevent ignition of Hazardous Atmospheres, do not open when
energized".
WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR
INTRINSIC SAFETY.
AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PUT
COMPROMETTRE, LA SECURITE INTRINSIQUE.
2210 Display Unit
Certificate: 1346165
Standards: CSA C22.2 No.0-10 (reaffirmed 2015),
CAN/CSA C22.2 60079-0:15,
CAN/CSA 60079-11:14,
CSA C22.2 No.142-M1987
Markings: Ex ib IIC T4 or Ex ib [ia] IIC T4, Ambient temperature
range [-40°C, +70°C] For installation refer to
drawing # 9150074-944.
WARNING: Substitution of components may
impair intrinsic safety
WARNING: POTENTIAL ELECTROSTATIC CHARGING
HAZARD.
U
I
o
Intrinsically
Safe Output
5.88 V
172.4 mA 0.253 W
Entity
parameters
111
I
P
C
o
o
o
μ
385.6 mA
0.678 W
8.8
F
μ
125.7 mA
0.726 W
0.14
F
I
P
C
i
i
i
μ
200 mA
1.3 W
0
F
P
C
o
o
o
μ
43
F
L
o
Intrinsically
0.17 mH
Safe Input
Entity
parameters
* Note: The associated apparatus shall be CSA certified with a
2.2 mH
maximum voltage rating Uo greater than 6.5 V or otherwise Ci
shall be considered 15.7 μF.
B.7 Europe
L
i
E1
ATEX Flameproof
0 mH
5600 Series Transmitter
Certificate: Sira 03ATEX1294X
Standards: EN 60079-0:2012/A11:2013, EN 60079-1:2014,
Markings:
Specific Conditions for Safe Use (X):
1. As light alloys may be used as the enclosure (or other parts)
they may be at the accessible surface of this equipment, in
the event of rare incidents, ignitions sources due to impact
and friction sparks could occur. This shall be considered
when the equipment is being installed in locations that
specifically require Group II, Category 1G equipment.
2. Under certain extreme circumstances, the non-metallic parts
of the equipment may be capable of generating an
ignition-capable level or electrostatic charge. Therefore,
L
o
when used for applications that specifically require Group II,
Category 1 equipment, the equipment shall not be installed
in a location where the external conditions are conductive to
mH
0.7
the build-up of electrostatic charge on such surfaces.
Additionally, the equipment non-metallic parts shall only be
cleaned with a damp cloth.
3. This equipment has flamepaths which differ from those in EN
60079-1 and are not intended for repair
Reference Manual
00809-0100-4024, Rev CA
U
I
P
i
i
i
12 V
400 mA
0.72 W
EN 60079-7:2015, EN 60079-11:2012,
EN 60079-26:2015 & EN 60079-31:2014
Non-IS Outputs (Primary, Secondary)
II ½ GD Ex db eb IIC T6 Ga/Gb (-40 °C to +70 °C)
Ex tb IIIC T 85 °C Db IP65
IS Outputs (Primary, Secondary and Display) alt.
with Non-IS Primary
Outputs
II (2) (1) ½ GD
Ex db eb [ib] [ia] IIC T6 Ga/Gb (-40 °C to +70 °C)
Ex tb IIIC T85 °C Db IP65
IS Outputs (Display) alt. with Non-IS Primary
Outputs
II (1) 1/2 GD
Ex db ib [ia] IIC T6 Ga/Gb (-40 °C to +70 °C)
Ex tb IIIC T85 °C Db IP65
C
L
i
i
n
mH
22
F*
0
Product Certifications

Advertisement

Table of Contents
loading

Table of Contents