4.3.1.6

Replacing the camera cable

IMPORTANT
Perform the replacement of the camera cable in a dust-free
environment.
Primescan
1. Attach the tool (A) (REF 6399229) to the cover nut (B) with the two
pins.
2. Release the cover nut (B) manually with the aid of the tool.
3. Remove the cover nut (B).
4. Take the camera cover (C) off the camera.
5. Remove the dust protection sticker (D). The sticker is no longer
required and can be disposed of.
6. Remove the spacer (E).
4.3 Replacing parts (without removing covers)