4 Repair
4.2 Removing covers and holders
58
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
4.2.2
Replacing the CEREC Primescan / Primescan scanner
holder
A CEREC Primescan /
Primescan scanner holder
cover parts set
C CEREC Primescan /
Primescan internal scan-
ner cable
ü The service cover is open (see "Opening the service cover [→ 52]").
ü The right cover is removed (see "Right cover [→ 54]").
1. Detach the scanner connector from the scanner cable and set the
scanner aside safely.
2. Unscrew the 3 screws (A) from the device and remove the scanner
holder laterally from the device.
3. Insert the new scanner holder and firmly screw the three screws
onto the device .
Dentsply Sirona
B CEREC Primescan / Primes-
can scanner holder
66 81 832 D3696
D3696.076.01.02.02 05.2019