Practical Example
Weighing in Components
Settings (changes in the factory settings required for this example):
Setup: App: Application 3: Formulation: Automatic storage: On, first value at stability
Setup: App: Application 3: Formulation: Minimum load for automatic storage: 100 digits
Setup: App: Application 3: Formulation: Evaluation mode, MR function: Final evaluation, print
Step
1.
Turn on the balance and configure
the settings as indicated above
2.
Delete old formulation data, if necessary
3.
Tare the balance
4.
Place the empty container
on the balance (here: 180.59 g)
5.
Tare the balance
6.
Weigh in the first component
(here: 42.88 g)
7.
Store components in the
formulation memory
Balance is tared automatically
Components are printed out
automatically
8.
Weigh in the next component
(here: 50.80 g)
Components are stored in the
totalization memory at stability
and printed out
Balance is tared automatically
9.
Repeat step 7 as required
10. Print final evaluation
(here: with total weight of all
components: 212.43 g)
11. Delete old formulation data, if necessary
92
Key (or instruction)
e
c
w
Place load on the balance
w
Place components
in container
M+
soft key
Place components
in container
MR
soft key
c
Display/Output
Comp2 +
50.80 g
--------------------
n
Tot.cp+
212.43 g
16.01.1997
--------------------
s
s
2
14:10