Configuration
Purpose
You can configure the balance to meet
individual requirements by entering user
data and setting parameters in the
Setup program.
The Setup menu contains the following
submenus:
– Balance functions
– Device parameters
– Application parameters
– Printout functions
– Device information
– Language
– Factory settings
Setting the Language
You can choose from 5 languages for
the information display:
– German
– English (factory setting)
– English with U.S. date | time format
– French
– Italian
– Spanish
Example: Setting the Language to "U.S. Mode"
Step
1. Select "Setup" menu
2. Select "Language"
and confirm
3. Select "U.S. mode"
4. Save language
5. Exit the Setup menu
12
Configuring the Balance for Use
in Legal Metrology
Set the menu access switch as described
below to configure the following
functions for use of the balance in legal
metrology:
– Display: Verification scale interval:
lower limit of the weighing capacity:
Min
– External calibration blocked
Preparation:
Remove the cap from the back of the
balance housing
Move the switch (1) upwards
Press key (or follow instructions)
o
Repeatedly
press
soft key,
q
then
soft key
O
soft key
q
soft key
l
soft key
oo
e
;
> When the switch is in the upper posi-
tion, the Setup menu is locked and the
balance can be used in legal metrology
When the switch is in the lower posi-
tion, the menu is accessible
> Note:
Do not move Switch 2