Example
Determining the weights of samples while monitoring the minimum sample quantity (here: SQmin: 30 mg)
Presettings (different from the factory settings):
Setup: Application parameters: Additional function (F4): SQmin
Step
1. Switch on the balance if necessary and
enter the presettings as shown above
2. Place the container into which the sample
will be filled onto the weighing pan and
tare the balance
3. Weigh a sample
(here: the minimum sample quantity
has not been reached)
4. Print out the weight
5. Weigh another sample
(here: the minimum sample quantity
has been exceeded)
6. Print out the weight
7. Display the minimum sample quantity
for 4 seconds
8. If necessary, weigh further samples
Press key(s) (or follow instructions) Display/Output
e
w
Place the sample on the
weihing pan
r
Place the sample on the
weighing pan
r
SQmin
soft key
*N
+
0.02 g
N
+ 16.38 g