Interface Description
Purpose
Your LA Reference balance comes
equipped with an interface port for
connection to a computer or other
peripheral devices.
You can use an on-line computer to
change, start and/or monitor the func-
tions of the balance and the application
programs. The interface port also has
four control lines for the over/under
checkweighing program. You can also
connect a hand switch or foot switch.
! Warning When Using Pre-wired
RS-232 Connecting Cables!
RS-232 cables purchased from other
manufacturers often have incorrect pin
assignements for use with Sartorius
balances. Be sure to check the pin
assignment against the chart below
before connecting the cable, and
disconnect any lines marked "Internally
Connected" (e.g., pin 6). Failure to do
so may damage or even completely ruin
your balance and/or peripheral device.
Available Features
Type of interface:
Operating mode:
Standard:
Transmission rates:
Parity:
Character format:
Handshake:
Operating mode:
Network address**:
Data output format of the balance:
* xBPI operating mode: 9,600 baud, 8 bits, odd parity, 1 stop bit
** Network address is only valid in the xBPI mode
Factory Settings:
Transmission rate:
Parity:
Stop bits:
Handshake:
Operating mode:
Network address:
Print manually/automatically:
Stop automatic printing:
Automatic printout, time-dependent: After 1 display update
Tare after indiv. printout:
Application initialization values:
Line format:
Preparation
See page 122 for the pin assignment chart
Serial interface
Full duplex
RS-232
150; 300; 600; 1,200; 2,400; 4,800; 9,600;
19,200 baud
Space, odd, even
1 start bit, 7-/8-bit ASCII, parity, 1 or 2 stop bits
2-wire interface: via software (XON/XOFF); 4-wire
interface: via hardware handshake lines (CTS/DTR)
SBI, xBPI*
0, 1, 2, ..., 30, 31
16 or 22 characters
1,200 baud
Odd
1 stop bit
Hardware
1 character after CTS
SBI
0
Manual after stability
Not possible
Off
Off
For other applications/GLP (22 characters)
117