Preparation
Example:
Set the parameters for calibration and adjustment; e.g., with manual calibration/adjustment, isoCAL off
Step
1. Switch on the balance,
if not already on
2. Select the Setup menu
3. Select "Balance/scale functions"
4. Select "Calibration/adjustment"
5. Select CAL key function
6. Select desired function and
confirm (e.g., "Internal cal./adj.")
7. Exit CAL key function
8. Select "Cal./adjustment sequence"
42
Press key(s) (or follow instructions) Display/Output