Xerox QLS-8100XE Operation & Setup Manual page 41

Quicklabel systems thermal transfer printer operation & setup guide
Table of Contents

Advertisement

6
Press the + or - key until Gap is displayed.
7
Press the SELECT key. The display will show the following prompt:
Place Liner @ Sensor
Then Press Select
8
Press and hold the FORM FEED/ADVANCE key until the liner is inside
the sensor bracket. On notched media, temporarily slide the media out so
that it is completely clear of the sensor assembly.
9
Press the SELECT key. The display will show the following prompt:
Place Media @ Sensor
Then Press Select
10 Press and hold the FORM FEED/ADVANCE key until the next label is
inside the Gap sensor.
11 Press the SELECT key. The display will show the following prompt:
Place Gap @ Arrow
Then Press Select
12 Press and hold the FORM FEED/ADVANCE key until the next gap is
positioned behind the Media Sensor Assembly and in close proximity to
the guide arrow indicated on the center plate.
13 Press the SELECT key and observe the gap size on the display. The gap
size should be the same as the measured gap of your media and the
following message will be displayed:
QLS-8100 Xe
Calibrate
Gap
Gap
Calibration OK
3-9