PG9602™ OPERATION AND MAINTENANCE MANUAL
In reassembly, the two sleeve retaining screws (5) MUST be reinstalled before
pressure is applied to the piston-cylinder module. Damage to equipment and injury to
personnel may result from pressurizing the piston-cylinder module without the sleeve
retaining screws installed.
To reassemble follow the steps above in reverse order. Follow the notes carefully and
refer to the lubrication recommendations in Section 5.3.4.
Figure 26. 50, 100 and 200 kPa/kg gas piston-cylinder modules (exploded view)
5.3.3
Cleaning Piston-Cylinders
There are two general methods for cleaning a piston-cylinder. They are a water / detergent
method, or a quicker method using Windex
•
Never
cylinder with your bare hands. Body oils and acids can permanently etch
the surfaces.
•
Do
familiarizing yourself with Section 5.3.1 of this manual.
disassembly may damage or destroy the piston-cylinder element.
© 2011 Fluke Calibration
touch the lapped surfaces (polished appearance) of the piston or
not
undertake
piston-cylinder
Page
Caution
®
brand (or equivalent) glass cleaner.
Caution
module
disassembly
172
without
Incorrect