1.2.1.1
AMH Automated Mass Handler (Optional)
Power Requirements:
Operating Temperature:
Dimensions:
Weight:
Power/Communications:
AMH Drive Air Supply:
AMH Vacuum Supply:
Pressure Connections:
Pressure:
CE conformance:
1.2.1.2
Embedded Features
•
Local control with 2 x 20 vacuum fluorescent display and 4 x 4 function
driven keypad.
•
Real time display and measurement of ambient (pressure, temperature,
humidity) and instrument (piston-cylinder temperature, piston position, piston
drop rate, piston rotation rate, piston rotation decay rate, reference vacuum)
conditions.
•
Real time mass-to-pressure and pressure-to-mass calculations taking into
consideration all environmental and operational variables.
•
Full gas and liquid fluid head corrections including DUT head correction and
piston position head correction.
atmosphere reference operation (gauge mode).
•
Adjustable mass loading resolution (0.01 g to 0.1 kg).
•
Audible prompts of instrument status (piston movement, Ready/Not Ready
indication) with override capability.
•
Integrated automated mass handling option (AMH-100-VAC).
•
Interfacing and automatic exploitation of external barometer via RS232.
•
Interfacing and automatic exploitation of any external vacuum gauge via
RS232.
•
Semi-automated differential mode to define low differential pressures at
various static pressures between vacuum and two atmospheres.
•
Storage and one step activation of metrological data on up to 17 piston-
cylinder modules, (3) mass sets and (3) mass loading bells.
•
Continuous pressure Ready/Not Ready indication based on measured
conditions.
•
Motorized, intelligent piston drive system based measured rotation rate with
operator alert and manual override.
•
Integrated automated pressure control with standard Fluke Calibration
pressure controllers.
•
Full RS232 and IEEE-488 communications with multi-level commands to set
and read all instrument functions.
15 VDC @ 2 A, 30 W max. consumption
15 to 35 ºC
41 cm H x 41 cm W x 36 cm D (16.3 in. x 16.1 in. x 14.1 in.)
12 kg (25 lbs)
Custom 8 pin connector
550 kPa (80 psi), ± 10%, minimal flow
At least 50 kPa (7.5 pi) under atmosphere, minimal flow
Quick connector DESO (double end shut off) type stem
Vacuum:
Quick connector SESO (single end shut off) type stem
Available, must be specified
Barometer head correction included for
Page 3
1. INTRODUCTION
© 2011 Fluke Calibration