4.3.3.1
If the optional AMH is being used, it can generate its own error messages during
operation. This usually occurs if the AMH is not properly setup, or has a
mechanical failure. The 3
indicate if the AMH has failed to operate as expected. If this occurs the
"AMHERR" command may be used to get the specific AMH error message. See
Section 3.7.2.8 for more details.
4.3.4

Command Descriptions

Each command description gives the full syntax showing usage. Ranges of parameters or
parameter types are indicated. There are two types of commands. The Common and Status
Commands support IEEE STD 488.2, while the PG9000 commands access all other functions.
4.3.4.1
PG9000 supports a set of commands that are common to all instruments
conforming to IEEE Std. 488.2 protocol. Though defined by the IEEE-488.2
standard, they also apply to PG9000 RS232 (COM1) communications. These
commands make it easy to perform basic functions for any device that
supports them. These command also cover the status reporting commands.
Refer to Section 4.4 for details on the status registers mentioned in these
commands. Query forms of these commands must be followed by a question
mark and IEEE-488.2 Common Commands always start with an asterisk ("*").
Unlike the other PG9000 commands, they must have a space instead of an
equals sign ("=") between the command and any arguments. Also unlike the
other PG9000 commands, if using the IEEE-488 port, the query form (command
is immediately followed by a "?") must be used to get a reply. If using the COM1
port and the command is not a query, "OK" will be returned.
∗CLS
Purpose
Syntax
Remarks
Example
© 2011 Fluke Calibration
REPLY
ERR #30
File end
ERR #35
PG9000 must first be setup for differential mode
ERR #36
Mass load invalid
ERR #37
External device invalid
ERR #38
External device configured incorrectly
ERR #39
External device reply invalid
ERR #40
Not ready
ERR #41
Measurement outside limits
ERR #42
Storage Fault
ERR #43
Storage verification error

AMH Errors

rd
character of the "PR" query reply will be an 'E' to

IEEE Std. 488.2 Common And Status Commands

Clear all of the status and event structures.
"∗CLS"
This program message clears the following evens and status registers:
Standard Byte Register (STB)
Standard Event Status Register (ESR)
Error Queue
AMH error message
Pending OPC operations
Command:
"*CLS"
Reply:
"OK" (using COM1. No reply if IEEE-488 port)