PG9602™ OPERATION AND MAINTENANCE MANUAL
Edit the average mass density of the
mass loading bell and press [ENT]. The
value defaults to the default density for
the typical mass loading bell for the
PG9000 model. The display is:
Edit the calibration report number (up to
ten digits) to the number of the current
calibration report of the mass loading bell
that is being added and press [ENT]. The
display is:
Edit the calibration report date to the date
of the current calibration report of the
mass loading bell that is being added and
press [ENT]. The display is:
Edit the true mass value of the mass
loading bell that is being added and press
[ENT]. The value defaults to the nominal
mass of the typical mass loading bell for the
PG9000 model. The display is:
Select <2yes> to save the all changes
made to the mass loading bell and exit.
Select <2no> to abandon all changes
made to the mass loading bell and exit.

3.12.1.12 Edit a Mass Loading Bell

To edit information contained in an existing mass loading bell file. Use this
function to change mass loading bell data after a recalibration.
See Section 3.12.1.11 describing adding a new mass loading bell. The principles
and procedures to add or edit a mass loading bell are identical except that the mass
loading bell to edit is selected from the existing mass loading bell screen.

3.12.1.13 View a Mass Loading Bell

To view information contained in an existing mass loading bell file.
To view information contained in an existing mass loading bell file press
[SPECIAL] and select <1PC/MS>, <3mass bell>, <1view>.
function is identical to the add function (see Section 3.12.1.11).
© 2011 Fluke Calibration
The AMH automated mass handler mass bell includes the
bell, the binary mass carrier and the mass lifting shaft.
PURPOSE
PRINCIPLE & OPERATION
Mass loading bell information is element specific metrological
data. Uninformed or accidental altering of this information may
lead to out of tolerance measurements.
information should only be edited by qualified personnel as part
of the PG9000 calibration or recalibration process.
PURPOSE
OPERATION
98
Page
Mass density?
5058 kg/m3
Note
Cal report number?
1775
Cal report date?
20040101
Bell #2832
True mass: 0.799985
Save bell S/N 2832?
1no 2yes
 Caution
Mass loading bell
The viewing