OPERATION
To select a SETUP file number to make it the active SETUP file, press [SETUP] and select
<1select>. In the following screen enter the number of the desired SETUP file and press
[ENT]. If a valid SETUP file number has been entered, operation returns to the previous run
screen with the selected SETUP file number displayed and active.
•
Entering an invalid SETUP file number causes an error message to be
displayed briefly and returns to the select SETUP file screen.
•
If the SETUP file selected specifies COM2 as the source of barometric
pressure, PG9000 will attempt to read a barometer on COM2 when
initializing the new SETUP file. If PG9000 is unable to read a barometer on
COM2, a communications time out message is displayed briefly and
operation returns to the select SETUP file screen. See Section 3.12.5.4 for
information on configuring COM2 to read an external barometer.
3.11.2
<2view>
PURPOSE
To view the variable source choices and values of any SETUP file number.
OPERATION
•
Selecting <2view> SETUP allows the variable source choices and values of
the active SETUP file for viewing only. <2view> does not make the viewed
SETUP file active. To select the active SETUP file, press [SETUP] and
select <1select>.
•
See Table 16 for a listing of the source choices for each variable.
To view an existing SETUP file, press [SETUP] and select <2view>. The display is:
1.
Entry field for number of SETUP file to be viewed. Defaults
to SETUP file currently selected. Must be a number from 2 to
10.
Enter the desired SETUP file number and press [ENT].
If a valid SETUP file number is entered, the display is:
<1atmP> to view the atmospheric pressure variable source choice and current value. The
value is displayed in the current pressure units (see Section 3.10.3).
<2ambT> to view the ambient temperature variable source choice and the current value.
The value is displayed in degrees Centigrade (°C).
<3%RH> to view the ambient relative humidity variable source choice and the current value.
The value is displayed in "%RH".
<4PCT> to view the piston-cylinder temperature variable source choice and the current value.
The value is displayed in degrees Centigrade (°C).
<5g> to view the gravity variable source choice and the current value. The value is displayed
in meters per second squared (m/s
Note
Note
2
).
Page 83
3. GENERAL OPERATION
View SETUP file:
#02
View #02 latmP 2ambT
3%RH 4PCT 5g 6READY
7vac
© 2011 Fluke Calibration