 OPERATION
To select a SETUP file number to make it the active SETUP file, press [SETUP] and select
<1select>. In the following screen enter the number of the desired SETUP file and press
[ENT]. If a valid SETUP file number has been entered, operation returns to the previous run
screen with the selected SETUP file number displayed and active.
Entering an invalid SETUP file number causes an error message to be
displayed briefly and returns to the select SETUP file screen.
If the SETUP file selected specifies COM2 as the source of barometric
pressure, PG9000 will attempt to read a barometer on COM2 when
initializing the new SETUP file. If PG9000 is unable to read a barometer on
COM2, a communications time out message is displayed briefly and
operation returns to the select SETUP file screen. See Section 3.12.5.4 for
information on configuring COM2 to read an external barometer.
3.11.2
<2view>
 PURPOSE
To view the variable source choices and values of any SETUP file number.
 OPERATION
Selecting <2view> SETUP allows the variable source choices and values of
the active SETUP file for viewing only. <2view> does not make the viewed
SETUP file active. To select the active SETUP file, press [SETUP] and
select <1select>.
See Table 16 for a listing of the source choices for each variable.
To view an existing SETUP file, press [SETUP] and select <2view>. The display is:
1.
Entry field for number of SETUP file to be viewed. Defaults
to SETUP file currently selected. Must be a number from 2 to
10.
Enter the desired SETUP file number and press [ENT].
If a valid SETUP file number is entered, the display is:
<1atmP> to view the atmospheric pressure variable source choice and current value. The
value is displayed in the current pressure units (see Section 3.10.3).
<2ambT> to view the ambient temperature variable source choice and the current value.
The value is displayed in degrees Centigrade (°C).
<3%RH> to view the ambient relative humidity variable source choice and the current value.
The value is displayed in "%RH".
<4PCT> to view the piston-cylinder temperature variable source choice and the current value.
The value is displayed in degrees Centigrade (°C).
<5g> to view the gravity variable source choice and the current value. The value is displayed
in meters per second squared (m/s
Note
Note
2
).
Page 83
3. GENERAL OPERATION
View SETUP file:
#02
View #02 latmP 2ambT
3%RH 4PCT 5g 6READY
7vac