Maintenance

Cleaning the Calibrator

To avoid personal injury or damage to the Calibrator, use
only the specified replacement parts and do not allow
water into the case.
To avoid damaging the case, do not use solvents or
abrasive cleaners.
Clean the Calibrator and pressure modules with a soft cloth dampened with water
or mild soap and water.

Replacing a Line Fuse

To avoid electrical shock hazard disconnect line power
before opening the case or line voltage selector.
The line power fuses and line voltage selector switch are located in a compartment
above the power switch on the right rear panel of the Calibrator. The fuse rating
label on the rear panel shows the correct replacement fuse for each line voltage
setting.
Table 6-1 lists the fuse part numbers for each line voltage setting. Figure 6-1
shows how to remove the fuse compartment cover.

Maintaining the Calibrator

Warning
Caution
Warning
Chapter 6
6-1