Optional Key Word - Fluke Norma 4000 User Manual

Guide for remote control for power analyzer
Hide thumbs Also See for Norma 4000:
Table of Contents

Advertisement

Example
:SYSTem
This key word denotes the :SYSTem command system. For commands of lower levels, the
complete path has to be specified, starting on the left with the highest level, the individual key
words being separated by a colon ":".
Example
INPut:COUPling AC
This command is at the second level of the INPut subsystem, see Figure 1-1. It selects
AC coupling of the input channel.
COUPling
Figure 1-1. Tree Structure of SCPI Command Systems Using INPut System as Example

Optional Key Word

Some command systems permit certain key words to be optionally inserted into the
header or omitted. These key words are marked in the description by square brackets. The
instrument must recognize the full command length for compatibility with the SCPI
standard. Some commands can be considerably shortened by omitting optional key words.
Example
INPut:FILTer[:STATe] ON
This command enables the antialiasing filter to be inserted into the signal path before it is
processed by SENSe subsystem. The following command has the same effect:
INPut:FILTer ON
An optional key word must not be omitted if its effect is specified in greater
detail by means of a numerical suffix.
INPut
GAIN
FILter
LPASs
FREQuency
Note
Remote Control Basics
Structure and Syntax of Device Messages
SHUNt
STATe
STATe
1
eya001.eps